Quick alignment tips
Dichotomy mode
- Adjust SEM to 100x magnification, 30kV and medium current
(spot size 3).
- Adjust SEM and optical focus
- Report current SEM magnification, set dwell time to 10µs
and scale to 8.
- Adjust optical settings until the e-beam is visible in the
optical image.
- Disable the auto brightness/contrast.
- Use “Dichotomy mode” to select smaller and smaller sub-areas
that move the e-beam closer to the center of the optical
image.
- Click on “move to center”.
- Select “Spot mode” to verify that the e-beam can be seen
in the optical image.
Spot mode
- Set binning of camera to 1
- Adjust optical settings and modify the brightness/contrast
until the e-beam is clearly visible.
- Double click on the center of the e-beam until it is in the
center of the optical image.
- Increase SEM magnification to desired level and repeat
alignment.
Fine alignment
- Ensure the SEM field of view is approximately the same as the
optical field of view.
- Report the SEM magnification.
- Click on “Fine alignment...” to perform automatic
scaling and sub-pixel alignment.